AMI Powercon
Program List and Explanations
5.2 Signal Outputs
5.2.1
5.2.1.1
5.2.1.2
5.2.1.3
As process
values
A-96.250.341 / 061015
Note:The navigation in the menu <Signal Output 1> and
<Signal Output 2> is equal. For reason of simplicity only the
menu numbers of Signal Output 1 are used in the following.
Signal Output 1: Assign process value, the current loop range and
a function to each signal output.
Parameter: Assign one of the process values to the signal output.
Available values:
Conductivity
Temperature
Sample flow
Cond. uc
Current Loop: Select the current range of the signal output.
Make sure the connected device works with the same current
range.
Available ranges: 0–20 mA or 4–20 mA
Function: Define if the signal output is used to transmit a process
value or to drive a control unit. Available functions are:
Linear, bilinear or logarithmic for process values.
See
As process values, p. 75
Control upwards or control downwards for controllers.
See
As control output, p. 77
The process value can be represented in 3 ways: linear, bilinear or
logarithmic. See graphs below.
[mA]
20
B
10
12
0 / 4
0.0
0.1
0.2
A
linear
B
bilinear
A
0.3
0.4
0.5
X Measured value
X
75