Relay 1 And 2 - Swann AMI Powercon Operator's Manual

Table of Contents

Advertisement

AMI Powercon
Installation
3.7.2
Relay
config.
Terminals
Normally
6/7: Relay 1
Open
8/9: Relay 2
Normally
6/7: Relay 1
Closed
8/9: Relay 2
32

Relay 1 and 2

Note:Max. load 1 A/250 VAC
Relay 1 and 2 can be configured as normally open or as normally
closed. Standard for both relays is normally open. To configure a
Relay as normally closed, set the jumper in the upper position.
NOTICE: Some error codes and the instrument status may
influence the status of the relays described below.
Jumper
pos.
Description
Inactive (opened) dur-
ing normal operation
and loss of power.
Active (closed) when a
programmed function is
executed.
Inactive (closed) during
normal operation and
loss of power.
Active (opened) when a
programmed function is
executed.
A
Jumper set as normally open (standard setting)
B
Jumper set as normally closed
For programming see menu Installation
tions, p.
70.
Relay configuration
0V
0V
A
B
Program List and Explana-
A-96.250.341 / 061015
6
7
6
7