AMI Powercon
Program List and Explanations
5.3.2.32.31
5.3.2.32.1
5.3.2.32.22
5.3.2.32.32
5.3.2.32.4
5.3.2.1
5.3.2.24
5.3.2.340
5.3.2.44
5.3.2.54
5.3.2.6
5.3.2.7
5.3.2.1
A-96.250.341 / 061015
Control Parameters
Range for each Parameter same as
Actuator = Motor valve
Dosing is controlled by the position of a motor driven mixing valve.
Run time: Time needed to open a completely closed valve
Range: 5–300 Sec.
Neutral zone: Minimal response time in % of the runtime. If the re-
quested dosing output is smaller than the response time, no
change will take place.
Range: 1–20 %
Control Parameters
Range for each Parameter same as
Function = Timer:
The relay will be activated repetitively depending on the pro-
grammed time scheme.
Mode: Operating mode (interval, daily, weekly)
Interval/Start time/Calendar: Dependent on options operating
mode.
Run time: time the relay stays active.
Range: 5–6'000 Sec
Delay: during run time plus the delay time the signal and control
outputs are held in the operating mode programmed below.
Range: 0–6'000 Sec
Signal Outputs: select the behavior of the signal outputs when the
relay closes. Available values: cont., hold, off
Output/Control: select the behavior of the control outputs when the
relay closes. Available values: cont., hold, off
Function = Fieldbus:
The relay will be switched via the Profibus input. No further param-
eters are needed.
5.2.1.43, p. 78
5.2.1.43, p. 78
83