Program List And Explanations; Messages; Diagnostics - Swann AMI Powercon Operator's Manual

Table of Contents

Advertisement

AMI Powercon

Program List and Explanations

9.

1 Messages

1.1 Pending Errors
1.1.5
1.2 Message List
1.2.1

2 Diagnostics

2.1 Identification
2.1.4
2.1.5
2.2 Sensors
2.2.1
2.2.1.4
2.2.1.5
70
Program List and Explanations
Provides the list of active errors with their status (active, acknowl-
edged). If an active error is acknowledged, the alarm relay is active
again. Cleared errors are moved to the Message list.
Shows the error history: Error code, date / time of issue and status
(active, acknowledged, cleared). 65 errors are memorized. Then
the oldest error is cleared to save the newest error (circular buffer).
In diagnostics mode, the values can only be viewed, not modified.
Desig.: Designation of the instrument.
Version: Firmware of instrument (e.g. V5.30-0912)
Factory Test: Test date of the Instrument, Motherboard and Fron-
tend
Operating Time: Years / Days / Hours / Minutes / Seconds
Cond. Sensor
Current value in µS
Raw value in µS
Cell Constant
QA History: Review the QA values (Number, Date-Time, Deviation
Conductivity, Deviation Temperature) of the last quality assurance
procedures. Only for diagnostic purpose. Max. 65 data records are
memorized.
Cal. History: Review diagnostic values of the last calibrations.Only
for diagnostic purpose.
Number; Date, Time
Cell constant
Max. 64 data records are memorized. One process calibration
corresponds to one data record.
A-96.250.341 / 061015