AMU Powercon
Instrument Setup
3.
3.1.
3.2.
Sensor
parameters
A-96.250.331 / 230320
Instrument Setup
Establish sample flow
1 Open flow regulating valve.
2 Wait until the flow cell has been completely filled.
3 Switch on power.
4 Adjust sample flow.
5 Let the instrument run-in for 1 h.
This recommendation is valid for rinsed cation exchanger
resin (nuclear grade) delivered by Swan.
Not rinsed cation exchanger resin from other suppliers may
require a run-in period of several hours to several days.
Programming
Program all sensor parameters in Menu 5.1.2.1,
<Installation> <Sensors> <Sensor parameters>:
The sensor characteristics are printed on the label of each sensor.
87-344.203
UP-Con1000SL
SW-xx-xx-xx
ZK = 0.0417
SWAN AG
DT = 0.06 °C
Enter the:
Cell constant [cm
Temperature correction [°C]
Cable length
Temperature compensation
Sensor type
Cell constant
Temperature correction
-1
]
23