AMI Powercon
Product Description
Fluidics
AMI Powercon
Specific
A-96.250.341 / 061015
The flow cell (QV-Flow) consists of the flow cell block [B], the flow
meter [C] and the flow regulating valve [E].
The conductivity sensor [A] with integrated temperature sensor is
screwed into the flow cell block [B].
The sample flows via the sample inlet [F] through the flow regulat-
ing valve [E], where the flow rate can be adjusted, into the flow cell
block [B], were the specific conductivity of the sample is measured.
The sample leaves the flow cell block via flow meter [C] through the
sample outlet [D].
A
Conductivity sensor
B
Flow cell block
C
Flow meter
A
B
C
D
E
F
D
Sample outlet
E
Flow regulating valve
F
Sample inlet