Sample Flow: Define at which sample flow an alarm should be is-
sued.
Flow Alarm: Program if the alarm relay should be activated if there
is a flow alarm. Choose between yes or no. The flow alarm will
always be indicated in the display, pending error list, saved in the
message list and the logger.
Available values: Yes or no
Note:Sufficient flow is essential for a correct measurement.
We recommend to program yes.
Alarm High: If the measuring values rises above the programmed
value E009 will be issued.
Range: 10–50 l/h
Alarm Low: If the measuring values falls below the programmed
value E010 will be issued.
Range: 0–9 l/h
Sample Temp.
Alarm High: If the measured value rises above the alarm high val-
ue, the alarm relay is activated and E007, is displayed in the mes-
sage list.
Range: 30–200 °C
Alarm Low: If the measured value falls below the alarm low value,
the alarm relay is activated and E008 is displayed in the message
list.
Range: -10 to + 20 °C
Case Temp. high
Alarm high: Set the alarm high value for temperature of electronics
housing. If the value rises above the programmed value E013 is is-
sued.
Range: 30–75 °C
Case Temp. low
Alarm low: Set the alarm low value for temperature of electronics
housing. If the value falls below the programmed value E014 is is-
sued.
Range: -10 to +20 °C