Program List and Explanations
5.3.2/3
5.3.2.1
5.3.2.20
5.3.2.300
5.3.2.400
A-96.250.341 / 061015
Relay 1 and 2: The contacts can be set as normally open or nor-
mally closed with a jumper. See
of relay contacts 1 or 2 is defined by the user.
Note:The navigation in the menu <Relay 1> and <Relay 2> is
equal. For reason of simplicity only the menu numbers of
Relay 1 are used in the following.
1 First select the functions as:
- Limit upper/lower,
- Control upwards/downwards,
- Timer
- Fieldbus
2 Then enter the necessary data depending on the selected func-
tion. The same values may also be entered in menu
Contacts, p. 73
Function = Limit upper/lower:
When the relays are used as upper or lower limit switches, program
the following:
Parameter: select a process value
Setpoint: If the measured value rises above respectively falls below
the set-point, the relay is activated.
Parameter
Range
0 S–300 mS
Conductivity
Temperature
-25 to +270 °C
Sample flow
0–50 l/h
0 S–300 mS
Cond. uc
Hysteresis: within the hysteresis range, the relay does not switch.
This prevents damage of relay contacts when the measured value
fluctuates around the alarm value.
Parameter
Range
0 S–300 mS
Conductivity
Temperature
0 to +100 °C
Sample flow
0–50 l/h
0 S–300 mS
Cond. uc