AMI Powercon
Maintenance
6.7.1
6.7.2
Assemble
A-96.250.341 / 061015

Exchange deaeration tube of cation exchanger bottle

1 Remove the inlet tube [C] to the pre-rinsed cation exchanger
bottle form the flowmeter [B].
2 Remove the flowmeter [B] from the flow cell [A].
3 Remove the sample collector [D] form the panel [I].
4 Unscrew and remove the tube fittings of the deaeration tubes
[E] and [F] from the sample connector.
5 Unscrew and remove the bottle holder [H] from the flow cell [A].
6 Unscrew and remove the tube fitting [G] from the bottle holder
[H].
7 Replace the 1 mm deaeration tube [F].
8 Screw the tube fitting into the bottle holder and tighten it.
9 Screw the bottle holder to the flow cell.
10 Screw the cation exchanger bottle into the bottle holder.
Do not tighten the bottle too firmly, this could damage the
gasket.
11 Before installing the sample collector [D] and the flowmeter [B]
replace the deaeration tube [E], see following chapter.

Exchange deaeration tube of pre-rinse bottle

1 Unscrew and remove the Flow regulating valve [K], with a
14 mm open-end wrench from the flow cell.
2 Unscrew and remove the tube Fitting from the bottle holder [J].
3 Replace the 2 mm deaeration tube [E].
4 Screw the flow regulating valve into the flow cell and tighten it
well.
1 Screw the sample connector onto the panel.
2 Screw the flow meter [B] onto the flow cell [A].
3 Connect the inlet tube [C] to the pre-rinsed cation exchanger
bottle with the elbow union of the flowmeter [B].