Electrical Requirements - Beckman Coulter Allegra 6 Series Instruction Manual

Table of Contents

Advertisement

Installation Requirements

ELECTRICAL REQUIREMENTS

2-2
• Select a location away from heat-producing laboratory equipment,
with sufficient ventilation to allow for heat dissipation.
• Position the centrifuge on a level surface (a sturdy table or labora-
tory bench for benchtop models) that can support the weight of the
centrifuge (refer to
SPECIFICATIONS
vibration.
• Allow a 7.6-cm (3-in.) clearance at each side and the back of the
centrifuge to ensure sufficient air circulation. Dimensions are
shown in Figure 2-1. The centrifuge must have adequate air venti-
lation to ensure compliance to local requirements for vapors
produced during operation.
• Do not operate the centrifuge in ambient temperatures exceeding
35°C (95°F). Relative humidity should not exceed 95%
(noncondensing).
Use the appropriate anti-rotation kit to secure the centrifuge to the
working surface. The anti-rotation kit is designed to minimize the
possibility of injury or damage that could result from centrifuge
movement in the event of a rotor mishap.
• The anchoring kit for a benchtop centrifuge contains bolts to
secure the centrifuge to the table or bench. Complete instructions
(SP-TB-007) for installing the anti-rotation kit are included with
the kit, which is shipped with the centrifuge.
• The anchoring kit for a kneewell centrifuge contains a stationary
tray to roll the centrifuge into for operation. Complete instructions
(GS6-TB-007) for installing the anti-rotation kit are included with
the kit, which is shipped with the centrifuge.
Allegra 6
120 VAC, 8 A, 60 Hz
230 VAC, 4 A, 50 Hz
100 VAC, 10 A, 50/60 Hz
A 1.8-m (6-ft) power cord with grounded plug is supplied with the
centrifuge. In regions where a different plug is required to meet local
electrical and safety requirements, contact your local Beckman
Coulter representative.
in Section 1) and resist
Allegra 6R /
Spinchron R
120 VAC, 12 A, 60 Hz
230 VAC, 8 A, 50 Hz
200 VAC, 10 A, 50/60 Hz
Allegra 6KR
120 VAC, 12 A, 60 Hz
230 VAC, 8 A, 50 Hz
200 VAC, 10 A, 50/60 Hz

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Spinchron rAllegra 6Allegra 6rAllegra 6kr

Table of Contents