ELECTRICAL REQUIREMENTS

• Ambient temperature during operation should not exceed
35°C (95°F). Relative humidity should not exceed 95%
(noncondensing).
!
CAUTION
Use the anti-rotation kit to secure the centrifuge
to the bench or table. The anti-rotation kit is
designed to minimize the possibility of injury or
damage that could result from centrifuge move-
ment in the event of a rotor mishap. Complete
instructions (SP-TB-007) for installing the anti-
rotation kit are included with the kit, which is
shipped with the centrifuge.
100-volt centrifuge. . . . . . . . . . . . . . . . 90–110 VAC, 12 A, 50/60 Hz
120-volt centrifuge. . . . . . . . . . . . . . . . . . 108–132 VAC, 10 A, 60 Hz
230-volt centrifuge. . . . . . . . . . . . . . . . . 207–253 VAC, 6.3 A, 50 Hz
Electrical supply . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Class I
Power cord . . . . . . . . . . . . . a 1.8-m (6-ft) power cord with grounded
NOTE
In regions where a different plug is required to
meet local electrical and safety requirements,
contact your local Beckman Coulter
representative.
Make sure the voltage imprinted on the name rating plate affixed to
the back of the centrifuge agrees with the line voltage of the outlet
used. With the power switch in the off (O) position, plug in both ends
of the centrifuge power cord. If there is any question about voltage,
have a qualified service person measure it under load while the drive
is operating.
To reduce the risk of electrical shock, this equipment uses a
three-wire electrical cord and plug to connect the centrifuge to
earth-ground. To preserve this safety feature:
• Make sure that the matching wall outlet receptacle is properly
wired and earth-grounded.
Installation Requirements
plug is supplied with the centrifuge
2-3