Set Inspection Item Text String Data
This command sets the text string data for the specified inspection item.
<Command Format>
Command
I
T
E
M
D
or
I
D
2
Inspection item
Command
number
(2 digits max.)
Space
(0×20)
<Response Format>
When the Command Is Processed Normally
O
K
CR
Delimiter
When the Command Is Not Processed Normally
E
R
CR
Delimiter
<Parameter Descriptions>
Inspection item number
External reference data
number
Text string data
FQ2-S4 User's Manual
Controlling Operation and Outputting Data with TCP No-protocol Communications
Inspection
item number
(2 digits max.)
A
T
A
2
Space
(0×20)
External reference
data number
(3 digits max.)
Space
Space
(0×20)
(0×20)
Specifies the inspection item number. (0 to 31)
Specifies the external reference data number. (0 to 999)
Refer to 12-2 External Reference Parameters p. 491 for details.
Specifies the text string data for the specified inspection item.
Refer to 12-2 External Reference Parameters p. 491 for details.
External reference
data number
(3 digits max.)
Space
Space
(0×20)
(0×20)
Text
CR
string data
Delimiter
Text
CR
string data
Delimiter