Ac Voltage, Dc Offset; Ac Voltage, Square Wave Characteristics; Ac Voltage, Triangle Wave Characteristics; Ac Current (Non-Sine Wave) - Fluke 5522A Operator's Manual

Multi-product calibrator
Hide thumbs Also See for 5522A:
Table of Contents

Advertisement

5522A
Operators Manual

AC Voltage, DC Offset

[1]
Range
3.3 to 32.999 mV
33 to 329.999 mV
0.33 to 3.29999 V
3.3 to 32.9999 V
9.3 to 92.999 mV
93 to 929.999 mV
0.93 to 9.29999 V
9.3 to 93.0000 V
6.6 to 65.999 mV
66 to 659.999 mV
0.66 to 6.59999 V
6.6 to 66.0000 V
[1]
Offsets are not allowed on ranges above the highest range shown above.
[2]
The maximum offset value is determined by the difference between the peak value of the selected voltage output and the
allowable maximum peak signal. For example, a 10 V p-p square wave output has a peak value of 5 V, allowing a maximum
offset up to
minimum outputs in each range.
[3]
For frequencies 0.01 to 10 Hz, and 500 kHz to 2 MHz, the offset uncertainty is 5 % of output, 1 % of the offset range.

AC Voltage, Square Wave Characteristics

Risetime @
Settling Time @
1 kHz
Typical
<10 S to 1 % of
<1 s

AC Voltage, Triangle Wave Characteristics

0.3 % of p-p value, from 10 % to 90 % point

AC Current (Non-Sine Wave)

Triangle Wave &
Truncated Sine Wave
Range
p-p
0.047 to
0.92999 mA
0.93 to
9.29999 mA
9.3 to
92.9999 mA
1-24
(Normal Channel)
Offset Range
0 to 500 mV
Triangle Waves and Truncated Sine Waves (p-p)
0 to 500 mV
0 to 500 mV
50 V to not exceed the 55 V maximum peak signal. The maximum offset values shown above are for the
Overshoot
@ 1 kHz
1 kHz Typical
Typical
<2 %
final value
Linearity to 1 kHz
Frequency
10 to 45 Hz
45 Hz to 1 kHz
[1]
1 to 10 kHz
10 to 45 Hz
45 Hz to 1 kHz
[1]
1 to 10 kHz
10 to 45 Hz
45 Hz to 1 kHz
[1]
1 to 10 kHz
Max Peak
[2]
Signal
Sine Waves (rms)
0 to 50 mV
80 mV
800 mV
0 to 5 V
8 V
0 to 50 V
55 V
0 to 50 mV
80 mV
800 mV
0 to 5 V
8 V
0 to 50 V
55 V
Square Waves (p-p)
0 to 50 mV
80 mV
800 mV
0 to 5 V
8 V
0 to 50 V
55 V
Duty Cycle Range
1 % to 99 % <3.3 V p-p.
0,01 Hz to 100 kHz
<1 % of p-p value, with amplitude >50 % of range
1-Year Absolute Uncertainty tcal 5 C
(% of output + % of range)
0.25 + 0.5
0.25 + 0.25
10 + 2
0.25 + 0.5
0.25 + 0.25
10 + 2
0.25 + 0.5
0.25 + 0.25
10 + 2
1-Year Absolute Uncertainty,
[3]
tcal 5 C
(% dc output + floor)
0.1 + 33 V
0.1 + 330 V
0.1 + 3300 V
0.1 + 33 mV
0.1 + 93 V
0.1 + 930 V
0.1 + 9300 V
0.1 + 93 mV
0.1 + 66 V
0.1 + 660 V
0.1 + 6600 V
0.1 + 66 mV
Duty Cycle Uncertainty
(0.02 % of period + 100 ns), 50 % duty cycle
(0.05 % of period + 100 ns), other duty cycles
from 10 % to 90 %
(0.8 % of period +100 ns)
Aberrations
Max Current
Resolution
Six digits
Six digits
Six digits