Oscilloscope Connections; How To Start The Sc1100 Option - Fluke 5522A Operator's Manual

Multi-product calibrator
Hide thumbs Also See for 5522A:
Table of Contents

Advertisement

5522A
Operators Manual
Oscilloscope Input Capacitance Measurement Specifications
Scope input selected
Measurement Range
Uncertainty
[1]
Measurement made within 30 minutes of capacitance zero reference. Scope option must be selected for at least five minutes prior
to any capacitance measurement, including the zero process.
Overload Measurement Specifications
Source Voltage
5 to 9 V

Oscilloscope Connections

Using the cable supplied with the SC1100 Option, connect the SCOPE output on the
Calibrator to one of the channel connectors on your oscilloscope (see Figure 10-1).
To use the external trigger, connect the TRIG OUT output on the Calibrator to the
external trigger connection on your oscilloscope. To use the external trigger and view its
signal with the calibration signal, connect the TRIG OUT output to another channel. See
your oscilloscope manual for details on connecting and viewing an external trigger.

How to Start the SC1100 Option

Press
appears in the Control Display. You can press any of the first four softkeys to go directly
to the VOLT, EDGE, LEVSINE, and MARKER calibration menus. Press the last softkey
to go to the OTHER menu (also shown below), allowing access to WAVEGEN, VIDEO,
PULSE, Impedance/Capacitance measurement (MEAS Z), and Overload (OVERLD)
menus. Press
describes each of these menus in detail.
10-8
Typical 'On' current
indication
100 to 180 mA
Figure 10-1. Oscilloscope Connection: Channel and External Trigger
(LED lit) to select the SC1100 Option. The SCOPE menu, shown below,
to return to the SCOPE menu from the OTHER menu. This manual
1 M
5 to 50 pF
(5 % of input + 0.5 pF)
Typical 'Off' current indication
10 mA
[1]
Maximum Time Limit DC or AC
(1 kHz)
setable 1 to 60 s
5522A
CALIBRATOR
gjh036.eps