Video Function Commands; Overload Function Commands - Fluke 5522A Operator's Manual

Multi-product calibrator
Hide thumbs Also See for 5522A:
Table of Contents

Advertisement

Video Function Commands

VIDEOFMT
(IEEE-488, RS-232, Sequential)
Selects the format for VIDEO mode.
Parameters: NTSC, PAL, PALM (for PAL-M), or SECAM
Example:
VIDEOFMT SECAM
VIDEOFMT?
(IEEE-488, RS-232, Sequential)
Returns the VIDEO mode format.
Parameters: None
Response:
NTSC, PAL, PALM (for PAL-M), or SECAM
VIDEOMARK
(IEEE-488, RS-232, Sequential)
Programs the VIDEO mode line marker location.
Parameters: Line marker number.
Example:
VIDEOMARK 10
VIDEOMARK?
(IEEE-488, RS-232, Sequential)
Returns the VIDEO mode line marker setting.
Parameters: None.
Response:
<character> SINE, SPIKE, SQUARE or SQ20PCT

Overload Function Commands

OL_TRIP?
(IEEE-488, RS-232, Sequential)
Returns the detected state of scope overload protection.
Parameters: None
Response:
Returns the number of seconds before protection was tripped. Returns 0 if
protection has not been tripped or if OVERLD mode not active.
TLIMIT
(IEEE-488, RS-232, Sequential)
Sets the OPERATE time limit for the OVERLD mode signal. The Calibrator
automatically returns to STANDBY if the UUT protection trips within this interval or at
the end of this interval if the protection has not tripped.
Parameters: 1 to 60 (seconds)
Example:
TLIMIT 30
SC600 Oscilloscope Calibration Option
Remote Commands and Queries
9
9-31