Ac Voltage (Non-Sine Wave) (Cont.); Ac Voltage, Dc Offset - Fluke 5502E Getting Started

Multi-product calibrator
Table of Contents

Advertisement

AC Voltage (Non-Sine Wave) (cont.)

Square Wave
Frequency
[1]
Range (p-p)
0.01 to 10 Hz
10 to 45 Hz
2.9 to 65.999 mV
45 Hz to 1 kHz
1 to 20 kHz
20 to 100 kHz
0.01 to 10 Hz
10 to 45 Hz
66 to 659.999 mV
45 Hz to 1 kHz
1 to 20 kHz
20 to 100 kHz
0.01 to 10 Hz
10 to 45 Hz
0.66 to 6.59999 V
45 Hz to 1 kHz
1 to 20 kHz
20 to 100 kHz
0.01 to 10 Hz
10 to 45 Hz
45 Hz to 1 kHz
1 to 20 kHz
6.6 to 66.0000 V
20 to 100 kHz
45 Hz to 1 kHz
1 to 10 kHz
[1]
To convert p-p to rms for square wave, multiply the p-p value by 0.5.
[2]
Uncertainty is stated in p-p. Amplitude is verified using an rms-responding DMM.

AC Voltage, DC Offset

[1]
Range
(Normal Channel)
3.3 to 32.999 mV
33 to 329.999 mV
0.33 to 3.29999 V
3.3 to 32.9999 V
9.3 to 92.999 mV
93 to 929.999 mV
0.93 to 9.29999 V
9.3 to 93.0000 V
6.6 to 65.999 mV
66 to 659.999 mV
0.66 to 6.59999 V
6.6 to 66.0000 V
[1]
Offsets are not allowed on ranges above the highest range shown above.
[2]
The maximum offset value is determined by the difference between the peak value of the selected voltage output and the
allowable maximum peak signal. For example, a 10 V p-p square wave output has a peak value of 5 V, allowing a maximum
offset up to ± 50 V to not exceed the 55 V maximum peak signal. The maximum offset values shown above are for the
minimum outputs in each range.
For frequencies 0.01 to 10 Hz, and 500 kHz to 2 MHz, the offset uncertainty is 5 % of output, ±1 % of the offset range.
[3]
1-Year Absolute Uncertainty, tcal
±5 °C, ±(% of output + % of range)
Normal Channel (Single Output Mode)
5.0 + 0.5
0.25 + 0.5
0.25 + 0.25
0.5 + 0.25
5.0 + 0.5
5.0 + 0.5
0.25 + 0.5
0.25 + 0.25
0.5 + 0.25
5.0 + 0.5
5.0 + 0.5
0.25 + 0.5
0.25 + 0.25
0.5 + 0.25
5.0 + 0.5
5.0 + 0.5
0.25 + 0.5
0.25 + 0.25
0.5 + 0.25
5.0 + 0.5
0.25 + 0.25
[3]
5.0 + 0.5
Max Peak
[2]
Offset Range
Sine Waves (rms)
0 to 50 mV
0 to 500 mV
0 to 5 V
0 to 50 V
Triangle Waves and Truncated Sine Waves (p-p)
0 to 50 mV
0 to 500 mV
0 to 5 V
0 to 50 V
Square Waves (p-p)
0 to 50 mV
0 to 500 mV
0 to 5 V
0 to 50 V
Max Voltage Resolution
[2]
Two digits on each range
Six digits on each range
Two digits on each range
Six digits on each range
Two digits on each range
Six digits on each range
Two digits on each range
Six digits on each range
1-Year Absolute Uncertainty, tcal
[3]
Signal
±5 °C
±(% of dc output + floor)
0.1 + 33 μV
80 mV
800 mV
0.1 + 330 μV
8 V
0.1 + 3300 μV
55 V
0.1 + 33 mV
0.1 + 93 μV
80 mV
800 mV
0.1 + 930 μV
0.1 + 9300 μV
8 V
55 V
0.1 + 93 mV
80 mV
0.1 + 66 μV
800 mV
0.1 + 660 μV
0.1 + 6600 μV
8 V
55 V
0.1 + 66 mV
Multi-Product Calibrator