Fluke 5522A Operator's Manual page 303

Multi-product calibrator
Hide thumbs Also See for 5522A:
Table of Contents

Advertisement

Table 9-10. SC600 Option Leveled Sine Wave Verification: Flatness (cont.)
Nominal Value
Frequency
(V p-p)
0.005
480 MHz
0.005
570 MHz
0.005
580 MHz
0.005
590 MHz
0.005
600 MHz
0.0075
50 kHz
0.0075
30 MHz
0.0075
70 MHz
0.0075
120 MHz
0.0075
290 MHz
0.0075
360 MHz
0.0075
390 MHz
0.0075
400 MHz
0.0075
480 MHz
0.0075
570 MHz
0.0075
580 MHz
0.0075
590 MHz
0.0075
600 MHz
0.0099
50 kHz
0.0099
30 MHz
0.0099
70 MHz
0.0099
120 MHz
0.0099
290 MHz
0.0099
360 MHz
0.0099
390 MHz
0.0099
400 MHz
0.0099
480 MHz
0.0099
570 MHz
0.0099
580 MHz
0.0099
590 MHz
0.0099
600 MHz
SC600 Oscilloscope Calibration Option
Measured Value
Deviation (V p-p)
(V p-p)
na
na
Verification Tables
1-Year Spec.
(V p-p)
0.0003
0.0003
0.0003
0.0003
0.0003
na
0.0002125
0.0002125
0.00025
0.00025
0.0004
0.0004
0.0004
0.0004
0.0004
0.0004
0.0004
0.0004
na
0.0002485
0.0002485
0.000298
0.000298
0.000496
0.000496
0.000496
0.000496
0.000496
0.000496
0.000496
0.000496
9
9-41