Ac Voltage, Dc Offset Specifications; Ac Voltage, Square Wave Characteristics; Ac Voltage, Triangle Wave Characteristics (Typical) - Fluke 5520A Manual

Hide thumbs Also See for 5520A:
Table of Contents

Advertisement

Manual Supplement

AC Voltage, DC Offset Specifications

[1]
Range
(Normal Channel)
3.3 to 32.999 mV
33 to 329.999 mV
0.33 to 3.29999 V
3.3 to 32.9999 V
9.3 to 92.999 mV
93 to 929.999 mV
0.93 to 9.29999 V
9.3 to 93.0000 V
6.6 to 65.999 mV
66 to 659.999 mV
0.66 to 6.59999 V
6.6 to 66.0000 V
[1]
Offsets are not allowed on ranges above the highest range shown above.
[2]
The maximum offset value is determined by the difference between the peak value of the selected voltage output and the
allowable maximum peak signal. For example, a 10 V p-p square wave output has a peak value of 5 V, allowing a maximum offset
up to ± 50 V to not exceed the 55 V maximum peak signal. The maximum offset values shown above are for the minimum outputs
in each range.
For frequencies 0.01 to 10 Hz, and 500 kHz to 2 MHz, the offset uncertainty is 5 % of output, ±1 % of the offset range.
[3]

AC Voltage, Square Wave Characteristics

Risetime
Settling Time
@ 1 kHz
@ 1 kHz
Typical
Typical
< 10 μs to 1 % of
< 1 μs
final value

AC Voltage, Triangle Wave Characteristics (typical)

Linearity to 1 kHz
0.3 % of p-p value, from 10 % to 90 % point
[2]
Offset Range
Sine Waves (rms)
0 to 50 mV
0 to 500 mV
800 mV
0 to 5 V
0 to 50 V
Triangle Waves and Truncated Sine Waves (p-p)
0 to 50 mV
0 to 500 mV
800 mV
0 to 5 V
0 to 50 V
Square Waves (p-p)
0 to 50 mV
0 to 500 mV
800 mV
0 to 5 V
0 to 50 V
Overshoot
Duty Cycle
@ 1 kHz
Range
Typical
1 % to 99 %,
< 3.3 V p-p,
< 2 %
0.01 Hz to 100
kHz
1-Year Absolute Offset
Max
Uncertainty, tcal ±5 °C
Peak
±(% of dc output + floor)
Signal
80 mV
0.1 + 33 μV
0.1 + 330 μV
8 V
0.1 + 3300 μV
0.1 + 33 mV
55 V
80 mV
0.1 + 93 μV
0.1 + 930 μV
8 V
0.1 + 9300 μV
55 V
0.1 + 93 mV
0.1 + 66 μV
80 mV
0.1 + 660 μV
0.1 + 6600 μV
8 V
55 V
0.1 + 66 mV
Duty Cycle Uncertainty
±(0.02 % of period + 100 ns), 50 % duty cycle
±(0.05 % of period + 100 ns), other duty cycles
from 10 % to 90 %
Aberrations
< 1 % of p-p value, with amplitude >50 % of range
20
5520A Operator
[3]
7/11

Advertisement

Table of Contents
loading

Table of Contents