Chapter 9, "Accessories.") A typical connection is shown in Figure 5-3.
IEEE-488
Complete the following procedure to test IEEE-488 operation using Win32 Interactive
Control.
1. Complete the "IEEE-488 Port Setup Procedure" earlier in this chapter to set up the
2. Connect the PC and Calibrator IEEE-488 ports using a standard IEEE-488 cable.
3. From the programs menu, select "NI-488.2M software for...(your operating system)".
4. From the NI488.2M software menu, select "Win32 interactive control".
5. A DOS window opens with a prompt as shown here:
[
6. At the prompt type the following line to activate the IEEE interface card:
7. The prompt reads <ud0:>. From this prompt type <ibwrt "remote"> then press the
5-8
Port
Controller
Figure 5-3. Testing the IEEE-488 Port
Calibrator for GPIB operation. Note the GPIB Address Port (default is 4).
(See Chapter 9, "Accessories," for IEEE-488 cables available from Fluke.)
<ibdev 0 4 0 10 1 0>
The second number in this line is the primary address of the calibrator. If the address
has been changed from the factory default, change this line accordingly.
ENTER (or RETURN) key.
IEEE-488 Cable
IEEE-488 Port
5522A Calibrator
UUT
gjh043.eps