Load the error mode and increment the selected edit field by .00001 mV.
x
x
IEEE-488
ISCE
(Instrument Status Change Enable command) Loads two bytes into the two 16-bit ISCE
mask registers (ISCE1 and ISCE0). (See "Instrument Status Change Enable Registers" in
Chapter 5 for more information.)
Parameter:
<value> (decimal equivalent of the 16 bits, 0 to 32767)
Example:
ISCE 6272
Load decimal 6272 (binary 0001010001000000) to enable bits 12 (SETTLED), 10
(REMOTE) and 6 (HIVOLT). This is equivalent to sending the commands
ISCE0 6272 and ISCE1 6272 (see below).
x
IEEE-488
ISCE?
(Instrument Status Change Enable query) Returns the two bytes from the two 16-bit
ISCE mask registers (ISCE1 and ISCE0). (See "Instrument Status Change Enable
Registers" in Chapter 5 for more information.)
Response:
<value> (decimal equivalent of the 16 bits, 0 to 32767)
Example:
ISCE? returns 6272
Returns decimal 6272 (binary 0001010001000000) if bits 12 (SETTLED), 10
(REMOTE), and 6 (HIVOLT) are set to 1.
x
IEEE-488
ISCE0
(Instrument Status 1 to 0 Change Enable command) Loads the two bytes into the 16-bit
ISCE0 register. (See "Instrument Status Change Enable Registers" in Chapter 5 for more
information.)
Parameter:
<value> (decimal equivalent of the 16 bits, 0 to 32767)
Example:
ISCE0 6272
Load decimal 6272 (binary 0001010001000000) to enable bits 12 (SETTLED), 10
(REMOTE) and 6 (HIVOLT).
x
IEEE-488
ISCE0?
(Instrument Status 1 to 0 Change Enable query) Returns the two bytes from the 16-bit
ISCE0 register. (See "Instrument Status Change Enable Registers" in Chapter 5 for more
information.)
Response:
<value> (decimal equivalent of the 16 bits, 0 to 32767)
Example:
ISCE0? returns 6272
Returns decimal 6272 (binary 0001010001000000) if bits 12 (SETTLED), 10
(REMOTE), and 6 (HIVOLT) are set to 1.
x
IEEE-488
ISCE1
(Instrument Status 0 to 1 Change Enable command) Loads the two bytes into the 16-bit
ISCE1 register. (See "Instrument Status Change Enable Registers" in Chapter 5 for more
information.)
Parameter:
<value> (decimal equivalent of the 16 bits, 0 to 32767)
x
RS-232
Sequential
x
x
RS-232
Sequential
x
x
RS-232
Sequential
x
x
RS-232
Sequential
x
x
RS-232
Sequential
x
x
Overlapped
Coupled
x
x
Overlapped
Coupled
x
x
Overlapped
Coupled
x
x
Overlapped
Coupled
x
x
Overlapped
Coupled
6
Remote Commands
Commands
6-21