"Adjusting the Phase" and "Synchronizing the Calibrator using 10 MHz
IN/OUT" later in this chapter.
How to Set Resistance Output
Complete the following procedure to set a synthesized resistance output at the Calibrator
front panel NORMAL terminals. If you make an entry error, press
display, then reenter the value.
1. Press
2. Connect the UUT as described earlier in this chapter under "Connecting the
Calibrator to a UUT."
Since this is a synthesized output, be sure the terminal connections from the
Calibrator to the UUT are LO to LO and HI to HI.
3. Set the UUT to measure resistance on the desired range.
4. Press the numeric keys and decimal point key to enter the desired resistance output
(maximum six numeric keys). For example, 12.3456.
5. Press a multiplier key, if necessary. For example, press
6. Press
.
7. The Control Display now shows the amplitude of your resistance entry. For example,
12.3456 k (below).
8. Press
into the Output Display (below is typical).
9. Press
The softkeys allow selection of three lead-compensation settings and ohms zero.
to clear any output from the Calibrator.
Note
. The calibrator clears your entry from the Control Display and copies it
to activate the calibrator output.
Front Panel Operation
How to Set Output
to clear the
.
nn096f.eps
nn097f.eps
nn098f.eps
4
4-33