How To Adjust The Phase - Fluke 5522A Operator's Manual

Multi-product calibrator
Hide thumbs Also See for 5522A:
Table of Contents

Advertisement

5522A
Operators Manual
3. Press the softkey FUNDMTL to select the Calibrator front panel terminals for the
4. Press the softkey HARMNIC to enter the desired harmonic (1 to 50), with a
5. Press

How to Adjust the Phase

When in the dual ac voltage and ac power output modes, you can set the calibrator to
source two signals with adjustable phase difference. All phase adjustments shift the AUX
waveform in relation to the NORMAL waveform. Phase shift adjustments are entered
either as degrees (0 to 180.00) or as a power factor (PF). A leading or positive phase
shift causes the AUX waveform to lead the NORMAL waveform; a lagging or negative
phase shift causes the AUX waveform to lag the NORMAL waveform.
The softkey PHASE is available after pressing the WAVE MENUS softkey that appears
when outputting dual ac voltages or ac power (shown below for ac power output)
4-44
fundamental output, either NORMAL or AUX. The harmonic appears on the AUX
terminals.
maximum frequency output of 10 kHz. For example, entering the 7th harmonic
(below). When the control display shows the desired value, press
one or more times to return to previous menus.
nn108f.eps
.
nn109f.eps