When calibrating thermocouples, it is especially important to use the correct hookup wire
and miniconnector between the Calibrator front panel TC terminal and the UUT. You
must use thermocouple wire and miniconnectors that match the type of thermocouple.
For example, if simulating a temperature output for a type K thermocouple, use type K
thermocouple wire and type K miniplugs for the hookup.
To connect the calibrator to a UUT, proceed as follows:
1. If the calibrator is turned on, press
terminals.
2. Make the connections to the UUT by selecting the appropriate figure from Table 4-1.
For capacitance outputs, null out stray capacitance by connecting the test leads to the
UUT, routing them (but not connecting) to the Calibrator on a non-conductive surface.
Null out the reading on the UUT using "rel," "offset," or "null," whichever method
applies, and then connect the test leads to the Calibrator.
5522A Output
Resistance
Capacitance
DC Voltage
AC Voltage
DC Current
AC Current
RTD Simulation
Thermocouple Simulation
Note:
See the discussion under "Four-Wire versus Two-Wire Connections" above.
to remove the output from the calibrator
Table 4-1. UUT Connections
4-2 Resistance - four-wire compensated
4-3 Resistance - two-wire compensated
4-4 Resistance - compensation off
4-5 Capacitance - two-wire compensated
4-6 Capacitance - compensation off
4-7 DC Voltage/AC Voltage
4-7 DC Voltage/AC Voltage
4-8 DC Current/AC Current
4-8 DC Current/AC Current
4-9 Temperature (RTD)
4-10 Temperature (Thermocouple)