AC Current (Sine Wave) (cont.)
Range
Frequency
10 to 100 Hz
29.00 to
329.99 A
100 Hz to 1 kHz
10 to 100 Hz
0.33 to
3.29999 mA
100 Hz to 1 kHz
10 to 100 Hz
3.3 to
32.9999 mA
100 Hz to 1 kHz
10 to 100 Hz
33 to
329.999 mA
100 Hz to 1 kHz
10 to 100 Hz
0.33 to
2.99999 A
100 to 440 Hz
10 to 100 Hz
[1]
3 to 20.5 A
100 Hz to 1 kHz
[1]
Duty Cycle: Currents <11 A may be provided continuously. For currents >11 A, see Figure B. The current may be provided 60-
T-I minutes any 60 minute period where T is the temperature in C (room temperature is about 23 C) and I is the output
current in Amps. For example, 17 A, at 23 C could be provided for 60-17-23 = 20 minutes each hour. When the 5522A is
outputting currents between 5 and 11 amps for long periods, the internal self-heating reduces the duty cycle. Under those
conditions, the allowable "on" time indicated by the formula and Figure B is achieved only after the 5522A is outputting currents
<5 A for the "off" period first.
[2]
For currents >11 A, Floor specification is 4000 A within 30 seconds of selecting operate. For operating times >30 seconds,
the floor specification is 2000 A.
[3]
For currents >11 A, Floor specification is 1000 A within 30 seconds of selecting operate. For operating times >30 seconds,
the floor specification is 5000 A.
[4]
Subject to compliance voltages limits.
Range
0.029 to 0.32999 mA
0.33 to 3.29999 mA
3.3 to 32.9999 mA
33 to 329.999 mA
0.33 to 2.99999 A
3 to 20.5 A
[1]
Subject to specification adder for compliance voltages greater than 1 V rms.
Absolute Uncertainty,
tcal 5 C
(% of output + A)
90 days
1 year
LCOMP On
0.2 + 0.2
0.25 + 0.2
0.5 + 0.5
0.6 + 0.5
0.2 + 0.3
0.25 + 0.3
0.5 + 0.8
0.6 + 0.8
0.07 + 4
0.08 + 4
0.18 + 10
0.2 + 10
0.07 + 40
0.08 + 40
0.18 + 100
0.2 + 100
0.1 + 200
0.12 + 200
0.25 + 1000
0.3 + 1000
[2]
0.1 + 2000
0.12 + 2000
[3]
0.8 + 5000
1.0 + 5000
Resolution A
0.01
0.01
0.1
1
10
100
Introduction and Specifications
Detailed Specifications
Max Distortion &
Noise 10 Hz to
100 kHz BW
(% of output + floor)
0.1 + 1.0
0.05 + 1.0
0.15 + 1.5
0.06 + 1.5
0.15 + 5
0.05 + 5
0.15 + 50
0.05 + 50
0.2 + 500
0.25 + 500
[2]
0.1 + 0
[3]
0.5 + 0
Max Compliance Voltage V rms [1]
7
7
5
5
4
3
1
Max
Inductive
Load H
400
[4]
400
1-15

Advertisement

Table of Contents
loading