Table 19: Data Port Pinouts - Nellcor OxiMax N-550B Service Manual

Table of Contents

Advertisement

Connecting To The Data Port
N-550B
5. Select baud rate by pressing Adjust Up or Adjust Down button (2400,
9600, or 19200 [default]).
6. Press the SatSeconds Alarm Limit button to set the selection.
Data is transmitted in the RS-232 and RS-422 formats. RS-232 data can be
transmitted a maximum of 25 feet (7.62 meters). The RS-422 data can be
transmitted a maximum of 4,000 feet (1,219.2 meters). The pin outs for the
data port are listed in Table 19.
TXD+ and TXD- are the differential transmit data pair. RXD+ and RXD- are
the differential receive pair.

Table 19: Data Port Pinouts

Pin
1
RXD+ (RS-422 positive input)
2
RXD (RS-232 input)
3
TXD (RS-232 output)
4
TXD+ (RS-422 positive output) (-5 to -12 VDC with no audible
alarm, 5 to 12 VDC with audible alarm)
5
Signal Ground (isolated from Earth Ground)
6
NC (No connection)
7
Normally Open (N.O. with no audible alarm), dry contact for Nurse
Call
8
Normally Closed (N.C. with no audible alarm), dry contact for
Nurse Call)
9
RXD- (RS-422 negative output)
Data Port Interface Protocol