TS54TDR PRO Test Set
Users Guide

Using the TDR

The TDR lets you measure the length of cables and the
distance to the first open or short on cables as long as
3,000 ft (914 m). See ""Time Domain Reflectometry
(TDR) Technology" on page 23 for information on how
the TDR operates.
The TDR operates only on cables that have two
conductors.
To use the TDR:
Optional: Change the length units, turn on the
1
Line Verification function, and set the velocity of
propagation to the correct value for the cable. See
"How to Change the Length Units", "Line
Verification", and ""VOP" below.
Press
.
Q
2
Connect the test leads to the cable.
3
Press
.
t
4
Below are examples of TDR test results:
TDR
There is a short across the two
Results
conductors 1267 feet from the test
Short at
set.
1267 ft
TDR
The cable is 2433 feet long, or
Results
there is an open in one or both of
Open at
the conductors 2433 feet from the
2433 ft
test set.
TDR
There is data or voltage on the
Voltage
line. The TDR does not operate on
Detect
an active line.
Lockout
16
Note

How to Change the Length Units

The TDR shows length measurements in feet or meters.
To change the unit of measurement:
Press
.
Q
1
Press
, press
M
C
2
TDR
to
, then press
M
Press
to put the flashing cursor next to
D
3
Units
, then press
M
Press
or
to select feet (
L
R
4
To save your selection before you exit or go back
5
to the main menu, press
Saved
.
To go back to the previous menu, press
press
or
until
C
D
.
M

Line Verification

The Line Verification function does tests for data and
voltage on the line when you press
senses data or voltage, it gives these responses:
Data Detect LOCK OUT
the line.
If there is voltage on the line, but no dial tone, the
display shows a voltage measurement and the test
set dials the ANAC number.
Active Telephone Line
tone on the line and there is no ANAC number
stored in memory location 9.
If there is a dial tone on the line and an ANAC
number in memory location 9, the test set
automatically dials the ANAC number and shows
the telephone number for the line. This helps you
identify the line so that you can make it inactive
before you do the TDR test again. Or, the number
can show you that you are connected to the
wrong line.
If Line Verification is off, and there is data or voltage on
the line, the display shows
Lockout
.
to put the flashing cursor next
again.
.
ft
m
) or meters (
).
. The display shows
M
Q
Back
shows, then press
. If the test set
t
: There is data on
: There is a dial
TDR Voltage Detect
. Or,