5080A/SC
Users Manual
OUT_IMP(?) <value >
Description Programs the scope options output impedance.
Parameters <value> = Z50
Example
Query
Verification Tables
Before the 5080A/SC Option leaves the Fluke factory, it is verified to meet its
specifications at the following test points. The verification test points are provided here as
a guide when re-verification is desired.
Nominal Value (p-p)
5.0 mV
5.0 mV
5.0 mV
10.0 mV
20.0 mV
20.0 mV
50.0 mV
89.0 mV
89.0 mV
100.0 mV
200.0 mV
200.0 mV
500.0 mV
890.0 mV
890.0 mV
1.0 V
1.0 V
2.0 V
5.0 V
5.0 V
10.0 V
20
riešenia na presné meranie
Z1M
OUT_IMP Z1M
OUT_IMP?
Table 3. Voltage Function Verification: AC Voltage into a 1 MΩ ohm Load
Frequency
45 Hz
100 Hz
1 kHz
1 kHz
100 Hz
1 kHz
1 kHz
1 Hz
45 Hz
1 kHz
100 Hz
1 kHz
1 kHz
45 Hz
1 kHz
100 Hz
1 kHz
1 kHz
45 Hz
1 kHz
1 kHz
Programs the oscilloscope option's output to
50 Ω.
Programs the oscilloscope option's output to
1 MΩ
Sets the scope options output impedance to
1 MΩ.
Returns the scope options output impedance
(Z50 or Z1M).
Measured Value (p-p)
http://www.elso.sk
1-Year
Deviation (mV)
Spec. (mV)
0.2175
0.2175
0.2175
0.235
0.27
0.27
0.375
0.5115
0.5115
0.55
0.90
0.90
1.95
3.215
3.215
3.7
3.7
7.2
17.7
17.7
35.2