Hardware Information

Electrical Connections

Electrical Connections

Never use cables other than the ones supplied by Agilent Technologies to ensure
NOT E
proper functionality and compliance with safety or EMC regulations.
Agilent InfinityLab LC Series Multicolumn Thermostats User Manual
The CAN bus is a serial bus with high-speed data transfer. The two
connectors for the CAN bus are used for internal module data transfer and
synchronization.
The ERI/REMOTE connector may be used in combination with other analytical
instruments from Agilent Technologies if you want to use features such as
start, stop, common shutdown, prepare, and so on.
With the appropriate software, the LAN connector may be used to control the
module from a computer through a LAN connection. This connector is
activated and can be configured with the configuration switch.
With the appropriate software, the USB connector may be used to control the
module from a computer through a USB connection.
The power input socket accepts a line voltage of 100 – 240 VAC ± 10 % with a
line frequency of 50 or 60 Hz. Maximum power consumption varies by
module. There is no voltage selector on your module because the power
supply has wide-ranging capability. There are no externally accessible fuses
because automatic electronic fuses are implemented in the power supply.