Wrong Valve - Agilent Technologies InfinityLab LC Series User Manual

Hide thumbs Also See for InfinityLab LC Series:
Table of Contents

Advertisement

7
Error Information
MCT Error Messages

WRONG VALVE

Error ID: 32130
Valve not supported in this module.
Probable cause
1
For G7116B:
NOT E
For G7116B Bio:
For G7116A:
For G7116A Bio-inert:
Agilent InfinityLab LC Series Multicolumn Thermostats User Manual
A valve head has been installed which is not
supported by G7116A.
G4232D (5067-4240: 2-position/10-port Valve 1300 bar)
G4231C (5067-4241: 2-position/6-port Valve 1300 bar)
G4234A (5067-4284: 6-position/14-port Valve 800 bar)
G4234C (5067-4273: 6-position/14-port Valve 1300 bar)
G4239C (5067-4233: 8-position/18-port Valve 1300 bar)
G5641A (5067-6682: 2-position/10-port Valve 1300 bar, bio)
G4232C (5067-4283: 2-position/10-port Valve 800 bar)
G4231A (5067-4282: 2-position/6-port Valve 800 bar)
G4237A (5067-4279: 4-position/10-port Valve 800 bar)
G5632A (5067-4132: 2-position/10-port Valve 600 bar, bio-inert)
G5631A (5067-4148: 2-position/6-port Valve 600 bar, bio-inert)
G5639A (5067-4134: 4-position/10-port Valve 600 bar, bio-inert)
Suggested actions
Only valve heads up to 800 bar or with up to four
positions can be used in G7116A. Exchange the
valve head with a suitable one. For changing the
valve head follow the instructions Replace Valve
Heads.