Leak - Agilent Technologies InfinityLab LC Series User Manual

Hide thumbs Also See for InfinityLab LC Series:
Table of Contents

Advertisement

7
Error Information
General Error Messages

Leak

Error ID: 0064
A leak was detected in the module.
The signals from the two temperature sensors (leak sensor and board-mounted
temperature-compensation sensor) are used by the leak algorithm to determine
whether a leak is present. When a leak occurs, the leak sensor is cooled by the
solvent. This changes the resistance of the leak sensor which is sensed by the
leak sensor circuit on the main board.
Probable cause
1
2
3
4
Agilent InfinityLab LC Series Multicolumn Thermostats User Manual
Condensation.
Loose column fittings.
Broken capillary.
Leaking column-switching valve seal.
Suggested actions
Use a higher temperature setpoint.
Ensure all fittings are tight.
Exchange defective capillaries.
Exchange the valve seal.
110

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

G7116bG7116a

Table of Contents