7
Error Information
General Error Messages
General Error Messages
General error messages are generic to all G HPLC modules.
Timeout
Error ID: 0062
The timeout threshold was exceeded.
Probable cause
1
2
Agilent InfinityLab LC Series Multicolumn Thermostats User Manual
The analysis was completed successfully,
and the timeout function switched off the
module as requested.
A not-ready condition was present during a
sequence or multiple-injection run for a
period longer than the timeout threshold.
Suggested actions
Check the logbook for the occurrence and source
of a not-ready condition. Restart the analysis
where required.
Check the logbook for the occurrence and source
of a not-ready condition. Restart the analysis
where required.
107