Using the Module
Exchange a Column
Install a column
1
Connect the column inlet capillary to the column inlet
using the Quick Connect Fitting.
3
Close the lever to make a leak-tight connection.
5
Use a wrench to counter the column while tightening
the capillary fitting.
7
Configure the column. For further information see Agilent Information Center or the online help of your CDS.
Agilent InfinityLab LC Series Multicolumn Thermostats User Manual
2
Turn the column onto the Quick Connect Fitting until it
is finger-tight.
4
Connect the capillary to detector to the column outlet.
When using the Quick Turn Fitting, tighten the fitting
until it is finger-tight.
6
Position the column into the column holder clip(s).