Error Information
MCT Error Messages

PELTIER OVERCURRENT

Error ID: 32098 (left), 32099 (right)
Probable cause
1
2
3

LEFT PELTIER ERROR

Error ID: 32096
Left Peltier hardware is not operational.
Probable cause
1
2

RIGHT PELTIER ERROR

Error ID: 32097
Right Peltier hardware is not operational.
Probable cause
1
2
Agilent InfinityLab LC Series Multicolumn Thermostats User Manual
Defective sensor
Defective mainboard
Defective cable
Defect heater element
Defective mainboard
Defect heater element
Defective mainboard
Suggested actions
Please contact your Agilent service representa-
tive.
Please contact your Agilent service representa-
tive.
Please contact your Agilent service representa-
tive.
Suggested actions
Please contact your Agilent service representa-
tive.
Please contact your Agilent service representa-
tive.
Suggested actions
Please contact your Agilent service representa-
tive.
Please contact your Agilent service representa-
tive.