Maintenance

Correcting Leaks

Correcting Leaks

When
If a leakage has occurred at the heat exchanger or at the capillary connections or at the column
switching valve.
Tools required
Description
Tissue
Pipette
Wrench, 1/4 – 5/16 inch
(for capillary connections)
1 Remove the door.
2 Use a pipette and tissue to dry the leak sensor area.
3 Observe the capillary connections and the column switching valve for leaks
4 Reinstall the door.
Agilent InfinityLab LC Series Multicolumn Thermostats User Manual
and correct, if required.
131