Valve Failed; Valve Tag Violation - Agilent Technologies InfinityLab LC Series User Manual

Hide thumbs Also See for InfinityLab LC Series:
Table of Contents

Advertisement

7
Error Information
MCT Error Messages

VALVE FAILED

Error ID: 32031
Switching of the valve failed.
Probable cause
1
2

VALVE TAG VIOLATION

Error ID: 32032
Reading the valve tag failed.
Probable cause
1
Agilent InfinityLab LC Series Multicolumn Thermostats User Manual
Mechanical problems. Friction too high or
blockages on the valve drive's motor or on
the valve head.
Defective sensor on the valve drive motor
A valve head has been exchanged
(hot-plugged) while the valve drive was still
powered on.
Suggested actions
Check valve head for correct installation.
Try to identify the source of trouble by install-
ing a different valve head if possible.
Contact your Agilent service representative.
Check valve head for correct installation.
Try to identify the source of trouble (valve
head or drive) by installing a different valve
head if possible.
Contact your Agilent service representative.
Suggested actions
For changing the valve head follow the
instructions "Replace Valve Heads." It is
important to have the MCT switched off for
at least 10 s after a new valve head has been
installed.
Contact your Agilent service representative.