This device is intended for operation in the electromagnetic
environment given below. The user of the device should ensure that
it is used in such an environment.
Tests of Resistance
to Disruption
Conducted high
frequency disturbance
variables in
accordance with
IEC 61000-4-6
Radiated high
frequency disturbance
variables in
accordance with
IEC 61000-4-3
76
IEC 60601
Test Level
Agreement
3 V
eff
3 V
150 kHz
to 80 MHz
3 V/m
80 MHz
3 V/m
to 2.5 GHz
Level of
Electromagnetic Environment
Use portable and mobile radio
sets at a safe distance from the
device (including the leads) not
closer than the recommended
protective distance, which is
calculated according to the
equation suitable for the
transmission frequency.
Recommended protective
distance:
d = 1.2
eff
d = 1.2
d = 2.3
P
With
transmitter in watts (W) in
accordance with the manufacturer
information on the transmitter and
d as the recommended protective
distance in metres (m).
The field strength of stationary
radio transmitters is lower than the
level of the agreement
frequencies in accordance with an
on-site investigation
Disruptions are possible near
devices that carry the following
symbol:
P
P
80 MHz to 800 MHz
P
800 MHz to 2.5 GHz
as the rated power of the
a
at all