3
Insert the cassette containing the emergency image plate into the cassette slot
as shown below.
Make sure to insert the cassette with the hinge [1] at the top and the locking
mechanism [2] at the bottom.
The image plate will be digitized using the speed class, i.e. the sensitivity,
corresponding to the emergency button as defined during configuration.
When the CR 25.0 has treated the emergency image plate, it displays the operator
main screen. The
accompanied by default ID data. The image processing station will give the
emergency image priority over the other images in the image processing queue.
If you do not enter a cassette within 15 seconds after pressing the emer-
gency button, the CR 25.0 will quit the emergency button function and return
to the operator main screen.
To change the speed class corresponding to the emergency button, contact
your local service organization.
4
Remove the cassette from the cassette slot.
38
1
2
digital image data
are transmitted to the image processing station