This device is intended for operation in the electromagnetic
environment given below. The user of the device should ensure that
it is used in such an environment.
Resistance to
Jamming Test
Discharge of static
electricity in
accordance with
IEC 61000-4-2
Fast transient
electrical
disturbance
variables / bursts in
accordance with
IEC 61000-4-4
Impulse voltages
(surges) in
accordance with
IEC 61000-4-5
Voltage
breakthroughs,
short term
interruptions and
variations in the
voltage supplied in
accordance with
IEC 61000-4-11
Magnetic field at
the supply
frequency (50/60
Hz) in accordance
with IEC 61000-4-8
• REMARK : U
2312A EN 20040909
IEC 60601
Test Level
+ 6 kV contact
discharge
+ 8 kV air discharge
+ 2 kV for network
leads
+ 1 kV for entry and
outlet leads
+ 1 kV push-pull
voltage
+ 2 kV common
mode voltage
• < 5% U
(> 95%
r
breakthrough of
U
) for ½ period
r
• 40% U
(> 60%
r
breakthrough of
U
) for 5 periods
r
• 70% U
(30%
r
breakthrough of
U
) for 25 periods
r
• < 5% U
(95%
r
breakthrough of
U
) for 5 s
r
3 A/m
is the alternating current in the network before the application of the test level.
r
Level of
Agreement
+ 6 kV contact
discharge
+ 8 kV air discharge
+ 2 kV for network
leads
+ 1 kV for entry and
outlet leads
+ 1 kV push-pull
voltage
+ 2 kV common
mode voltage
• < 5% U
(> 95%
r
breakthrough of
U
) for ½ period
r
• 40% U
(> 60%
r
breakthrough of
U
) for 5 periods
r
• 70% U
(30%
r
breakthrough of
U
) for 25 periods
r
• < 5% U
(95%
r
breakthrough of
U
) for 5 s
r
3 A/m
Electromagnetic Environ-
ment Guidelines
Floors should consist of wood,
concrete or ceramic tiles. The
relative humidity must be at
least 30%, if the floor is made of
synthetic material.
The quality of the voltage
supplied should correspond to a
typical commercial or clinical
environment.
The quality of the voltage
supplied should correspond to
that of a typical commercial or
clinical environment.
The quality of the voltage supply
should correspond to that of a
typical commercial or clinical
environment.
If the user wants the device to
work continuously, even when
the energy supply is interrupted,
it is recommended to use an
energy supply free of
interruptions or a battery.
Magnetic field at the network
frequency should correspond to
the typical values as they are in
a commercial and clinical
environment.
75