Emergency buttons
Two emergency buttons are located at the front of the CR 25.0. The emergency
buttons determine the speed class, i.e. the sensitivity, which will be used to digitize
the image plate. The sensitivity associated with the emergency buttons has been
set during the configuration of your system. For more information, contact your
local service organization.
The emergency buttons have the following labels:
For digitizing images of the trunk.
For digitizing images of the limbs.
In the interchangeable configuration, the emergency buttons allow you to
digitize emergency image plates without ID data.
In the dedicated configuration, the emergency buttons allow you to digitize an
image plate while you are entering the identification data via the ID Station.
Audio signals
The CR 25.0 gives status information via beeps. The length of the beep indicates
the response of the system to a key command.
• A short beep means that CR 25.0 has accepted the key command and is starting
the operation.
• A long beep means that you have pressed a non-active key or that the CR 25.0
has rejected the key command.
• An interval beep accompanies an error, locked or warning message. Refer to
'Troubleshooting checklist'
2312A EN 20040909
on page 54.