Agilent Technologies E1437A User Manual page 172

20 msample/second adc with filters and fifo
Table of Contents

Advertisement

E1437A
ASCII Overview and Commands
TRIGGER:PHASE:CAPTURE
Prepares for LO phase capture in frequency-synchronized, multiple-module
zoom measurements.
Command Syntax:
TRIGGER:PHASE:CAPTURE
Example Statements:
OUTPUT 70924;"trigger:phase:Capture
Description:
Use this function if you intend to subsequently use TRIGGER:DELAY:ACTUAL? to
capture the LO phase on the next SYNC assertion. You should send
TRIGGER:DELAY:CAPTURE to only one module in the system (typically the
master) after you have completed all frequency and filter setup functions, since
those functions take the module out of the phase_capture mode. Therefore, you
should call TRIGGER:DELAY:CAPTURE just prior to starting the measurement.
When the FREQUENCY:SYNC mode is turned off, the TRIGGER:DELAY:CAPTURE
function is not needed because the module will revert to the phase:capture mode by
default.
See Also:
TRIGGER:PHASE:ACTUAL?, TRIGGER_SETUP
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
TRIGGER:PHASE:CAPTURE
command
command
5-49