Agilent Technologies E1437A User Manual page 150

20 msample/second adc with filters and fifo
Table of Contents

Advertisement

E1437A
ASCII Overview and Commands
Parameter Definitions:
0. . The state of this parameter does not affect any transition where freq ≠0.
Whether the real or complex data is saved and ultimately sent to the output port is
determined by the DATA:TYPE command.
parameter
parameter
sample frequency. freq is the desired center frequency divided by the ADC sample
frequency. For example, selecting .25 with a sample clock frequency of 20 MHz will
yield a center frequency of 5.0 MHz. The ADC sample frequency is returned by the
CLOCK:FS? command. Negative frequencies select the negative image of the signal,
which is spectrally inverted from the input signal.
Comments:
Although the freq parameter is a double floating point number, its effective
resolution is 1/(1024*10^9) or 20 µHz when fs=20.48 MHz. The actual frequency will
be set to the nearest available value. This value is returned by the
FREQUENCY:CENTER? command. In multi-module systems this value represents
the pending value rather than the current value when a frequency change is
incomplete due to a pending SYNC signal transition.
In multiple-module systems it is often desirable to force the frequency change to
occur synchronously in order to preserve the phase relationship of the LOs. This is
accomplished by setting the sync parameter to ON for all the modules which are to
be changed. See the first example below.
In configurations involving synchronous operation of multiple E1437 modules, the
FREQUENCY:SETUP command provides a mechanism to force all LOs to the same
phase. This can be done by first setting the frequency to zero. See the second
example below.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
selects either a phase continuous or phase reset transition when the freq =
value
0
OFF causes phase to be reset to zero when combined with a frequency change to zero
ON combined with a frequency change to zero does not reset the phase, thereby
1
generating a complex DC measurement at baseband.
controls when a frequency transition is implemented.
value
0
OFF allows an immediate frequency change.
ON. In multiple-module systems, setting this parameter ON prepares the modules for a
frequency change, but does not actually bring about the change until the next ADC clock
corresponding to the next assertion of the shared SYNC signal. The SYNC transition is
1
generated by calling the MEAS:CONTROL command. Note that returning sync to OFF
before the SYNC signal transition has occurred forces an immediate asynchronous
frequency change.
is a number between -0.5 and +0.5, which will be interpreted as a fraction of the
cmplxdc parameter definition
sync parameter definition
FREQUENCY:SETUP
command/query
5-27