E1437A
ASCII Overview and Commands
LBUS:MODE
Set and query local bus mode.
Command Syntax:
LBUS:MODE <mode>
mode::=012 3
Query Syntax:
LBUS:MODE?
Example Statements:
OUTPUT 70924;"Lbus:Mode 2"
Description:
LBUS:MODE sets the local bus to either generate, append, insert or pipeline data.
The data port must be set to the local bus with the DATA:PORT command before
these modes take effect.
Parameter Definitions:
DATA:PORT command. The state of this parameter is unaffected by switching back
and forth between the local bus and the VME backplane with the DATA:PORT
command.
parameter
Reset Values:
lbusMode=PIPELINE
See Also:
DATA:PORT
5-36
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
selects the transmission mode of the local bus when it is enabled by the
value
PIPELINE causes the E1437 to pipe data through from modules on its left without
0
appending or inserting its own data.
GENERATE forces the module addressed to generate data only, not passing through data
1
from other modules on the local bus
APPEND causes the E1437 to pass through data from modules on its left and append its
2
data to the end
INSERT causes the E1437 to place its data on the local bus and then pass through data
3
from modules on its left.
Module(s) to Left
mode parameter definition
HP E1437
GENERATE
INSERT
APPEND
PIPELINE
LBUS:MODE
command/query
command/query
Module to Right