Agilent Technologies E1437A User Manual page 132

20 msample/second adc with filters and fifo
Table of Contents

Advertisement

E1437A
ASCII Overview and Commands
CLOCK:SETUP
Sets all timing parameters. This description also includes information on the
following commands which set or query the timing parameters individually:
CLOCK:DSP selects the clock used to drive the decimation/zoom section.
CLOCK:FS provides the frequency of an external sample clock.
CLOCK:MASTER determines whether a module shares its ADC clock.
CLOCK:MULTI:SYNC specifies whether the module uses a shared clock and sync.
CLOCK:SOURCE selects the source of the ADC clock.
Command syntax:
CLOCK:SETUP <multisync>,<source>,<dsp>,<master>,<fs>
multisync::= 012
source::= 0123
dsp::= 01
master::= 012
fs <numeric>
CLOCK:MULTI:SYNC 012
CLOCK:SOURCE 0123
CLOCK:DSP 01
CLOCK:MASTER 012
CLOCK:FS <numeric>
Query syntax:
CLOCK:DSP?
CLOCK:FS?
CLOCK:MASTER?
CLOCK:MULTI:SYNC?
CLOCK:SOURCE?
Example Statement:
OUTPUT 70924;"Clock:setup 1,2,0,2,10000000"
OUTPUT 70924;"Clock:Multi:Sync 2"
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
numeric::=>0-20600000
numeric::=100000-20600000
CLOCK:SETUP
command/query
command/query
5-9

Advertisement

Table of Contents
loading

Table of Contents