20 msample/second adc with filters and fifo
Table of Contents

Advertisement

E1437A
ASCII Overview and Commands
INPUT:RANGE:AUTO
Performs auto-ranging.
Command Syntax:
INPUT:RANGE:AUTO <sec>
sec::=<numeric>
Description:
This command sets the range of a E1437 to the lowest value that will not cause an
ADC overload to occur. The algorithm will start at the lowest range and move up
until there is no ADC overload.
Parameter definitions:
detected. Setting this parameter to 0.0 will result in this time being set automatically
according to an algorithm that depends on block size and filter bandwidth.
NOTE
An autorange that is pending or in progress will be aborted if an INPUT:RANGE
or another INPUT:RANGE:AUTO command is received.
See Also
INPUT:SETUP
5-30
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
numeric::=≥0 seconds
is the time in seconds to take data at each range to insure that an overload is