ASCII Overview and Commands
FREQUENCY:CENTER:RAW
Provides a fast way to set the center frequency.
Command Syntax:
FREQUENCY:CENTER:RAW <coarse>, <fine>
coarse::=0 to 2047
fine::=0 to 499999999
Query Syntax:
FREQUENCY:CENTER:RAW?
Example Statements:
OUTPUT 70924;"FREQUENCY:CENTER:RAW 1024,1000000
Description:
This command sets the center frequency without relying on the internal E1437
microprocessor to do any floating point computations, since the internal
microprocessor does not have a floating point co-processor. The resulting center
frequency is:
fs*((coarse/2048)+(fine/1.024*10^12))
Parameter Definitions:
See Also:
FREQUENCY:SETUP, CLOCK:FS:GET, DATA:TYPE, MEAS:CONTROL
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
sets high frequencies or a low resolution frequency component.
sets very low frequencies or a high resolution frequency component.