Chapter 4: Verifying the System
Method 3: Measuring the Noise Floor

Preparing the Hardware

Do This
1 Connect the system as shown
in
Figure 4-5 on page
calibrated 50Ω load fitted to the RF input.
The Agilent receiver requires a 30 minute
warm-up period to meet the guaranteed
specifications.
2 Turn on the Agilent receiver.
4-14 Wireless Solutions Getting Started Guide
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Additional Information
There is a three minute delay after power-up or
4-13, using a
re-configuration, before the internal temperature sensor
is activated and measurements are nominally correct.
For more information on how to connect the receiver and
laptop, refer to step 1 and step 2 of any of the
procedures listed in
Setup"
Chapter 2, "Receiver-Based System