Chapter 4: Verifying the System
Method 1: Using a Signal Generator

Testing the Signal

Do This
1 In the Markers group, click Add and To Max. The level and frequency are displayed at the top-left
2 In the Band group, select Downlink.
3 In the Freq/Channel group, change the
Center frequency according to the
4 On the signal generator, set the frequency
according to the
Table 4-2 Signal Generator Downlink Settings
Model
GSM1900, TDMA and CDMA PCS band systems
GSM DCS1800 band systems
GSM900 band systems
Cellular CDMA and TDMA band systems
NOTE
Wait for at least one minute before using the displayed marker value. The
default trace display is set to display a running average of 32. With time, the
marker value will approach the value of the signal generator output.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Additional Information
corner of the Spectrum display.
Table
4-2.
You should see a signal trace showing a single
Table
4-2.
spectral peak in the middle of the display.
Figure 4-3 on page 4-9
Frequency and Power Settings
1959 MHz at -60 dBm
1843 MHz at -60 dBm
943 MHz at –60 dBm
882 MHz at -60 dBm
Wireless Solutions Getting Started Guide 4-7
shows a example trace.