Chapter 4: Verifying the System
Method 1: Using a Signal Generator

Preparing the Software

Do This
1 Start the application program by going to the
Start menu, selecting Programs, Agilent E74xx.
2 Create a new project called
Verification Test 1 (External generator).
3 Click Configuration and then add the receiver
hardware.
4 Click Collection and select the Spectrum VFP
(Virtual Front Panel)
5 Click Maximize button in the upper-right corner
of the main window.
6 Connect the signal generator output to the RF
input of the receiver.
7 Turn on the signal generator.
8 Set the measurement software as follows:
Band: Uplink
Frequency: See
Span: 1 MHz
Averaging: Running
Averages: 32
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Additional Information
The software starts.
We recommend you use the automatic
configuration wizard for this process. Refer to
Chapter 2, "Automatic
Refer to
The display size increases to fill the available
space.
You should see a signal trace showing a single
spectral peak in the middle of the display.
Figure 4-2 on page 4-6
Table 4-1 on page 4-6
Wireless Solutions Getting Started Guide 4-5
Configuration".
Chapter 2, "Hardware Installation Tips"
shows a example trace.