Safety Instrumented Systems (4-20 mA only)
April 2019
8.8.2
Special tools
Not required.
8.8.3
Product repair
The Rosemount 5300 Level Transmitter is repairable by major component replacement.
All failures detected by the transmitter diagnostics or by the proof-test must be reported.
Feedback can be submitted electronically at
(Contact Us).
8.8.4
Reset to factory settings
Transmitters with option code QS or QT are shipped with a special factory pre-
configuration. If the device needs to be reset to factory settings, please contact your local
Emerson representative to get the factory pre-configuration file to be uploaded to the
device or use the backup file created upon receiving the transmitter. This will ensure the
proper settings for your safety device are configured and your QS or QT certificate is still
valid.
8.9

Specifications

For general specifications, see
5300 Level Transmitter
The Rosemount 5300 Level Transmitter must be operated in accordance with the
functional and performance specifications provided in
Reference conditions can be found in
8.9.1
SIS reference
Failure rate data
The
FMEDA report
regarding failure rate analysis.
Failure values
Safety accuracy: 2.0%
Safety response time: at least below 8 seconds for specific configurations
Higher Safety Response Time is allowed for other user-selectable configurations, for
example damping, level rate etc. See
Self-diagnostics test interval: at least every 90 minute
(9)
The safety accuracy of the Rosemount 5300 Level Transmitter safety-certified option is ±2% of full span (± 0.32 mA).
268
Specifications and reference data
Product Data
Sheet.
Reference
includes failure rate data, assessment details, and assumptions
(9)
Set alarm limits
Go.EmersonProcess.com/Contact-Us
or refer to the Rosemount
Specifications and reference
conditions.
for more information.