Lcd Error Messages - Emerson 5301HxxxxxxxxxxxxxxZZ Series Reference Manual

Guided wave radar level transmitter
Table of Contents

Advertisement

Reference Manual
00809-0100-4530
Note
Analog output status messages may not appear on the integral display while the
Rosemount 5300 Level Transmitter runs in HART 7 multidrop mode.
Table 7-7: Analog Output Status
Message
Not connected
Alarm Mode
Saturated
Multidrop
Fixed Current mode
PV out of Limits
Span Too Small
Invalid Limits
7.3

LCD error messages

Figure 7-1: The Rosemount 5300 Display Panel Displaying an Error Message
Table 7-8: Error Messages Displayed on the Rosemount 5300 Display Panel
Error message
RAM FAIL
Reference Manual
Description
Analog output hardware is not
connected.
The analog output is in Alarm Mode.
The analog output signal value is
saturated, i.e. equal to the saturation
value.
The transmitter is in Multidrop Mode. The
analog output is fixed at 4 mA.
The analog output is in fixed current
mode.
The Primary variable is out of range.
The configured span is too small.
The given Upper and Lower Sensor Limits
are invalid.
Service and troubleshooting
Action
Check Error and Warning Messages to
find the reason for the Alarm.
No action needed.
This is the normal setting when a device is
used in Multidrop configuration.
This mode is used when calibrating the
Analog Output channel.
Check the Upper and Lower Range Value
Check the Upper and Lower Range Value.
Check that the difference between the
Upper and Lower Sensor Limits is greater
than the Minimum Span.
Description
An error in the gauge data memory (RAM) has been
detected during the startup tests.
Note
This resets the gauge automatically.
April 2019
177

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents